1.0 Purpose : To provide an instruction for the validation of QC Chemist.
2.0 Objective : To provide a documented procedure for the validation of QC Chemist.
3.0 Scope : This procedure is applicable for QC Chemist of ----Pharmaceuticals Ltd,.
4.0 Responsibility :
· Primary : QC Chemist / QC-Officer
· Secondary : Overall: QC – In charge
5.0 Procedure :
· Following method used for the validation of QC Chemist to confirmed his / her analytical skills.
Ø QC In-charge identifies a previously analyzed and approved batch sample.
Ø Identified batch sample shall be coded and hand over to QC Chemist with all the
details of test procedure.
· QC Chemist qualified on the criterion as follows:
Ø QC Chemist performs the analysis and documents the result as per Good Laboratory Practices.
Ø QC Chemist performs the analysis and meets the acceptance criteria as per below
mentioned table.
Ø The QC Chemist working in laboratory shall be qualified on one or more tests as
required by the QC In-charge.
Ø The QC Chemist shall be qualified once a year.
Ø A record shall be maintained.
Ø Incase of QC Chemist fails to meet the criteria, the QC Chemist shall be trained and re-qualified. The training record shall be recorded
Acceptance Criteria for Chemist Validation
Sr. No. | Test | Acceptance Criteria |
01 | Assay test by HPLC | % RSD of Retention Time and Area of six replicate injection should not more than 2 % |
02 | Impurity test by HPLC | % RSD of two replicate injection for highest impurity should not more than 10 % |
03 | IR test | Three separate analysis, which shall be concordant with each other. |
04 | Water / LOD test | % RSD of triplicate analysis should not more than 2 % 10 % moisture / LOD range 0 to 0.5 % 5 % moisture / LOD range 0.5 to 2.0 % 2 % moisture / LOD range 2 % above |
05 | UV test | % RSD of three replicated absorption of solution of particular concentration at specified wavelength should not be more than 2 % |
06 | GC test | % RSD of Retention Time and Area of six replicate injection should not more than 2 % |
07 | Sp. Gravity test | % RSD of six replicates reading should not more than 2 % |
08 | Specific Optical Rotation test | % RSD of six replicates reading should not more than 2 % |
09 | Assay by chemically test | % RSD of six replicate analysis should not be more than 2 % |
No comments:
Post a Comment