Instrument : Gas Chromatograph
Make : Shimadzu
Model : GC-2010
Date of calibration :
Next calibration due on :
A. UPKEEP OF INSTRUMENT:
Instrument is clean and free from dust
CHROMATOGRAPHIC CONDITIONS
Parameters | Test condition |
Column | Rtx 624 (30m × 0.53mm, 3µm) or equivalent |
Oven temp. program | 40 ºC-3min-@30 ºC/min-220 ºC-6 min |
Detector Temperature | 220 ºC |
Injector Temperature | 240 ºC |
Head Pressure | 20 Kpa |
Linear velocity | 31.0 cm/sec |
Injection Volume | 1 ml Gaseous phase |
Equilibration Temperature | 105 ºC |
Transfer line Temperature | 110 ºC |
Equilibration Time | 20 minutes |
Run Time | 15 minutes |
1 : 10 |
I) (a) Retention Time repeatability :
Run # | Observed Retention Time (RT) | Acceptance criteria | |
Iso Propyl Alcohol (IPA) | Tetrahydrofuran (THF) | ||
Run-1 | (% RSD: NMT 1.5 %) | ||
Run-2 | |||
Run-3 | |||
Run-4 | |||
Run-5 | |||
Mean | |||
% RSD |
Conclusion:
(b) Area Repeatability:
Run # | Iso Propyl Alcohol (IPA) | Tetrahydrofuran (THF) | Acceptance criteria |
Run-1 | (% RSD: NMT 15 %) | ||
Run-2 | |||
Run-3 | |||
Run-4 | |||
Run-5 | |||
Mean | |||
% RSD |
II) Linearity :
Make triplicate injection of each injection volume. Draw the linearity curve as concentration vs. area.
For Isopropyl alcohol (IPA)
Area | Concentrations(ppm) | ||||
500 | 400 | 300 | 200 | 100 | |
Area 1 | |||||
Area 2 | |||||
Area 3 | |||||
Mean |
Result:
Coefficient of correlation : | Limit: Not less than 0.99 |
Conclusion:
For Tetrahydrofuran (THF)
Area | Concentrations(ppm) | ||||
72.0 | 57.5 | 43.1 | 28.7 | 14.4 | |
Area 1 | |||||
Area 2 | |||||
Area 3 | |||||
Mean |
Result:
Coefficient of correlation: | Limit: Not less than 0.99 |
II) Temperature calibration for GC
Following parts will be calibrated at the temperature points indicated.
a) Column Oven Temperature :
Set Temp. (ºC) | Measured Temp.(ºC) | Limit(ºC) | Remarks |
50 | ±3.2 | ||
100 | ±4.2 | ||
200 | ±5.5 | ||
300 | ±5.5 |
b) Injector Port Temperature:
Set Temp. (ºC) | Measured Temp.(ºC) | Limit (ºC) | Remarks |
100 | ±4.2 | ||
200 | ±5.5 | ||
300 | ±5.5 |
c) Detector Port Temperature:
Set Temp. (ºC) | Measured Temp. (ºC) | Limit (ºC) | Remarks |
100 | ±4.2 | ||
200 | ±5.5 | ||
300 | ±5.5 |
d) HT 200HT Incubator Temperature:
Set Temp. (ºC) | Measured Temp. (ºC) | Limit (ºC) | Remarks |
50 | ±3.2 | ||
100 | ±4.2 | ||
150 | ±5.5 |
Carrier gas flow calibration:
Flow will be calibrated using digital flow meter at room temperature.
a) Column flow for packed column:
Set column flow in ml/min | Measured column flow in ml/min | Limit in ml/min | Acceptance criteria |
20.0 | 18.0 to 22.0 | For 20 and 30ml/min limit is ±10% For 50ml/min limit is±20% | |
30.0 | 27.0 to 33.0 | ||
50.0 | 40.0 to 60.0 |
Conclusion:
b) Column flow for capillary column:
Set column flow in ml/min | Measured column flow in ml/min | Limit in ml/min | Acceptance criteria |
2.5 | 2.25 to 2.75 | ±10% | |
5.0 | 4.5 to 5.5 |
Conclusion:
c) Split flow for capillary column:
Set split ratio | Set column flow in ml/min | Total flow in ml/min | Measured flow in ml/min | Limit of flow in ml/min | Acceptance criteria |
1:5 | 2.0ml/min | 10.0ml/min | 8.0 to 12.0 | ±20% | |
1:10 | 10.0ml/min | 100.0ml/min | 80.0 to 120.0 |
Conclusion :
Calibration Status: Satisfactorily/ Not Satisfactoril
Remark: Hence the instrument stands/does not stands calibrated w.r.t.
1. Detector Precision.
2. Detector linearity.
3. Temperature Calibration.
Flow Calibration